Anschluss (Druckluft):
4 / 6 mm tube connector
Anschluss (Probeneinlass):
12 mm barbed stem for 12 mm ID hose
Austauschbare Sensoren:
ISE, reference electrode, PT100
Breite:
482 mm
Druckbereich:
0,5 - 6 bar
Druckluft:
4 - 7 bar filtered and dried air
4 - 7 bar filtered and dried air
Durchflussrate:
40 - 300 L/h
Gehäuse-Schutzklasse:
IP65
Genauigkeit:
± 2 ± 3 %
Gewicht:
25 kg
25 kg approximately
Höhe (Panel):
753 mm
Kalibriermethode:
Manual 2-point or 1-point calibration method (automatic calibration method to be ordered separately)
Measuring Principle 2:
Automatic titration using potentiometric endpoint detection based on volumetric analysis
Messbereich:
Fluoride 0.1 - 500 mg/L F
Messintervall:
5 - 999 min frei programmierbar
Methode:
Measuring Fluoride Ion Selective Electrode (ISE), type 8470B
Niveauregulierung:
sample, reagents, calibration solution and chemical cleaning
Optionen:
External dilution, automatic calibration, automatic heating device, chemical cleaning, conditioning, reagent tray, free-standing cabinet, wall mount
Probenahmemodus:
Fixed or loop
Probenahmequalität:
No filtration required, only in coarse particle separation
Probenströme:
1
Probentemperatur:
5 - 50 °C (41 - 122 °F)
Probenvolumen:
200 - 1000 mL per cycle
Process connection (drain):
2x Kabel-Durchführung für 2 - 6 mm Durchmesser
Prozessanschluss (Abfluss):
3 x cable glands for diameter 13 - 18 mm
Prozessanschluss (Spülwasser):
6 mm barbed stem for 6 mm ID hose
Pumpen:
peristaltic reagent pump for sample conditioning (additional peristaltic or piston pumps to be ordered separately)
Reagent capacitiy:
10 L canisters supplied with instrument
Relais:
1 collective alarm; warning or alarm system
2 high or low limits on the concentrations
Rinse water pressure:
0,5 - 6 bar
Sensoren:
Fluoride Ion Selective Electrode (ISE) type 8470B, reference electrode type 8483B, temperature sensor PT100
Spannungsversorgung (Hz):
60 Hz
Spannungsversorgung (Volt):
110 V AC
Stromausgang:
2 x 0/4 - 20 mA linear, logarithmic or
bilinear programmable
Tiefe:
122 mm
Umgebungstemperatur:
5 - 55 °C
Wartungs-Intervall:
2 to 4 weeks application specific
Wartungs-Vorgaben:
1 to 4 weeks application specific